{"id":1582163781,"title":"HANNA HI2300-02 EC\/TDS\/NaCl\/°C Meter - Autorange","handle":"hi2300-02","description":"\u003cp style=\"text-align: justify;\"\u003e\u003cstrong\u003eHI 2300\u003c\/strong\u003e measures EC. TDS. NaCl and temperature. In conductivity and TDS ranges (up to 500 mS\/cm and 400 g\/L respectively) the instrument automatically chooses the best scale to maintain the highest accuracy.\u003c\/p\u003e\n\u003cp style=\"text-align: justify;\"\u003eEC calibration is a one-point procedure. Selectable calibration points are 0.00 µS. 84.0 µS. 1413 µS. 5.00 mS. 12.88 mS. 80.0 mS. and 111.8 mS selected according with the expected measurement range. NaCl calibration is a one-point procedure at 100.0% NaCl. Use HI 7037L calibration solution as a 100% NaCl standard solution. This instrument utilizes a four ring potentiometric probe with platinum sensors to offer versatility over typical amperometric designs. By utilizing the four ring method. it is possible to measure very low or high conductivity levels without changing probes. Three options of compensating for temperature are available for this instrument:        \u003c\/p\u003e\n\u003cp style=\"text-align: justify;\"\u003eAutomatic (ATC): The EC probe has a built-in temperature sensor which is used to automatically compensate the EC\/TDS reading (from -9.9°C to 120.0°C). using the selected reference temperature (20 or 25ºC) and temperature compensation coefficient from (0.0 to 6.0%)\/ºC\u003c\/p\u003e\n\u003cp style=\"text-align: justify;\"\u003eManual (MTC): The temperature value. shown on the secondary LCD. can be manually set with the ARROW keys. The compensation is referenced at the selected temperature. All the other parameters of temperature compensation are settable similar to ATC.\u003c\/p\u003e\n\u003cp style=\"text-align: justify;\"\u003eNo Compensation (NoTC): For actual conductivity or TDS measurement. the temperature value shown on the secondary LCD is not taken into account.\u003c\/p\u003e\n\u003cp style=\"text-align: justify;\"\u003eCell constant is selectable between 0.5 and 1.700. TDS factor is selectable between 0.40 and 0.80.\u003c\/p\u003e\n\u003cp style=\"text-align: justify;\"\u003eThe HI 2300 also provides users with GLP capabilities. Good Laboratory Practice (GLP) is a set of functions that allows storage and retrieval of data regarding the status of the system.\u003c\/p\u003e\n\u003cp style=\"text-align: justify;\"\u003eAfter a successful calibration. the meter automatically stores the date and time of calibration. the calibration solution used and the resulting cell constant value. All this information can be later recalled by the user. Other features include a lock range function and stability indicator.\u003c\/p\u003e\n\u003cp style=\"text-align: justify;\"\u003eFor PC communication. use the optional HI 92000 software and HI 920013 USB cable. The software is provided with an exclusive online guide of all the commands available and allows data printing. plotting and exporting.\u003c\/p\u003e\n\u003cp style=\"text-align: justify;\"\u003e \u003c\/p\u003e\n\u003cp style=\"text-align: justify;\"\u003e\u003cstrong\u003eOrder Information:\u003c\/strong\u003e   HI 2300-02 (230V) is supplied with HI 76310 conductivity probe. 12 VDC adapter and instructions.\u003c\/p\u003e\n\u003cdiv\u003e\u003cstrong\u003eProduct Manuals\u003c\/strong\u003e\u003c\/div\u003e\n\u003cp\u003e\u003cimg width=\"20\" height=\"22\" align=\"absmiddle\" alt=\"\" src=\"http:\/\/www.hannainst.com\/usa\/images\/icon-pdf.gif\" border=\"0\"\u003e  Manual: \u003ca title=\"Download Product Manual\" href=\"http:\/\/www.hannainst.com\/manuals\/man2300Y_09_12.pdf\" target=\"_blank\" rel=\"noopener noreferrer\"\u003eDownload\u003c\/a\u003e\u003c\/p\u003e\n\u003cp\u003e\u003cstrong\u003eSpecification :\u003c\/strong\u003e\u003c\/p\u003e\n\u003ctable class=\"spec_table\"\u003e\n\u003ctbody\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eRange\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e\u003cstrong\u003e\u003csmall\u003eEC\u003c\/small\u003e\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e0.00 to 29.99 µS\/cm; 30.0 to 299.9 µS\/cm; 300 to 2999 µS\/cm; 3.00 to 29.99 mS\/cm; 30.0 to 200.0 mS\/cm; up to 500.0 mS\/cm (actual EC)\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003cstrong\u003e\u003csmall\u003eTDS\u003c\/small\u003e\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e0.00 to 14.99 mg\/L (ppm); 15.0 to 149.9 mg\/L (ppm); 150 to 1499 mg\/L (ppm); 1.50 to 14.99 g\/L (ppt); 15.0 to 100.0 g\/L (ppt); up to 400.0 g\/L (actual TDS). with 0.80 conversion factor\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003cstrong\u003e\u003csmall\u003eNaCl\u003c\/small\u003e\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e0.0 to 400.0%\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003cstrong\u003e\u003csmall\u003eTemperature\u003c\/small\u003e\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e-20.0 to 120.0°C\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eResolution\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e\u003cstrong\u003e\u003csmall\u003eEC\u003c\/small\u003e\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e0.01 µS\/cm; 0.1 µS\/cm; 1 µS\/cm; 0.01 mS\/cm; 0.1 mS\/cm \u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003cstrong\u003e\u003csmall\u003eTDS\u003c\/small\u003e\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e0.01 mg\/L; 0.1 mg\/L; 1 mg\/L; 0.01 g\/L; 0.1 g\/L \u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003cstrong\u003e\u003csmall\u003eNaCl\u003c\/small\u003e\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e0.1%\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003cstrong\u003e\u003csmall\u003eTemperature\u003c\/small\u003e\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e0.1°C\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eAccuracy @ 25°C\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e\u003cstrong\u003e\u003csmall\u003eEC\u003c\/small\u003e\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e±1% of reading ± (0.05 µS\/cm or 1 digit)\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003cstrong\u003e\u003csmall\u003eTDS\u003c\/small\u003e\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e±1% of reading ± (0.03 mg\/L or 1 digit)\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003cstrong\u003e\u003csmall\u003eNaCl\u003c\/small\u003e\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e±1% of reading\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003cstrong\u003e\u003csmall\u003eTemperature\u003c\/small\u003e\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e±0.4°C\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eCalibration\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e\u003cstrong\u003e\u003csmall\u003eEC\u003c\/small\u003e\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003eautomatic. one point with six memorized values (84. 1413. 5000. 12880. 80000. 111800 µS\/cm) \u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003cstrong\u003e\u003csmall\u003eNaCl\u003c\/small\u003e\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003eone point. with HI 7037 calibration solution\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003cstrong\u003e\u003csmall\u003eTemperature\u003c\/small\u003e\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003etwo point. at 0 and 50°C\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eTemperature Compensation\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003eautomatic or manual from -20.0 to 120.0°C\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eTemperature Coefficient\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003eselectable from 0.00 to 6.00%\/°C (EC and TDS only)\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eTDS Factor\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003eselectable from 0.40 to 0.80 (default value: 0.50)\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eProbe\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003eHI 76310 platinum. four ring conductivity\/TDS probe with internal temperature sensor and 1 m (3.3') cable (included)\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003ePC Connection\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003eopto-isolated USB\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eLogging\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003elog on demand. 500 samples\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eAuto-off\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003eafter five minutes of non-use (can be disabled)\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003ePower Supply\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e12 VDC adapter (included)\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eEnvironment\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e0 to 50°C (32 to 122°F); RH max 95%\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eDimensions\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e235 x 218 x 108 mm (9.2 x 8.5 x 4.2”)\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eWeight\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e 1.3 kg (2.9 lbs.)\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003c\/tbody\u003e\n\u003c\/table\u003e\n\u003cp\u003e\u003cstrong\u003e \u003c\/strong\u003e\u003c\/p\u003e","published_at":"2017-11-22T12:35:41+13:00","created_at":"2015-09-01T15:15:27+12:00","vendor":"Hanna Instruments","type":"Electrochemical","tags":["Manufacturer_Hanna Instruments","Type_Conductivity\/TDS Meters","Type_pH \u0026 ORP Meters"],"price":57000,"price_min":57000,"price_max":57000,"available":true,"price_varies":false,"compare_at_price":118700,"compare_at_price_min":118700,"compare_at_price_max":118700,"compare_at_price_varies":false,"variants":[{"id":4857292101,"title":"Default Title","option1":"Default Title","option2":null,"option3":null,"sku":"HI2300-02","requires_shipping":false,"taxable":true,"featured_image":null,"available":true,"name":"HANNA HI2300-02 EC\/TDS\/NaCl\/°C Meter - Autorange","public_title":null,"options":["Default Title"],"price":57000,"weight":0,"compare_at_price":118700,"inventory_quantity":1,"inventory_management":null,"inventory_policy":"deny","barcode":null,"requires_selling_plan":false,"selling_plan_allocations":[]}],"images":["\/\/acornsci.co.nz\/cdn\/shop\/products\/ec-tds-metre-de-laboratoire-avec-gamme-nacl-230_1-1_2015.jpeg?v=1465536556"],"featured_image":"\/\/acornsci.co.nz\/cdn\/shop\/products\/ec-tds-metre-de-laboratoire-avec-gamme-nacl-230_1-1_2015.jpeg?v=1465536556","options":["Title"],"media":[{"alt":"HI 2300-02 EC\/TDS\/NaCl\/°C Meter - Autorange - Acorn Scientific","id":21298708579,"position":1,"preview_image":{"aspect_ratio":1.0,"height":1000,"width":1000,"src":"\/\/acornsci.co.nz\/cdn\/shop\/products\/ec-tds-metre-de-laboratoire-avec-gamme-nacl-230_1-1_2015.jpeg?v=1465536556"},"aspect_ratio":1.0,"height":1000,"media_type":"image","src":"\/\/acornsci.co.nz\/cdn\/shop\/products\/ec-tds-metre-de-laboratoire-avec-gamme-nacl-230_1-1_2015.jpeg?v=1465536556","width":1000}],"requires_selling_plan":false,"selling_plan_groups":[],"content":"\u003cp style=\"text-align: justify;\"\u003e\u003cstrong\u003eHI 2300\u003c\/strong\u003e measures EC. TDS. NaCl and temperature. In conductivity and TDS ranges (up to 500 mS\/cm and 400 g\/L respectively) the instrument automatically chooses the best scale to maintain the highest accuracy.\u003c\/p\u003e\n\u003cp style=\"text-align: justify;\"\u003eEC calibration is a one-point procedure. Selectable calibration points are 0.00 µS. 84.0 µS. 1413 µS. 5.00 mS. 12.88 mS. 80.0 mS. and 111.8 mS selected according with the expected measurement range. NaCl calibration is a one-point procedure at 100.0% NaCl. Use HI 7037L calibration solution as a 100% NaCl standard solution. This instrument utilizes a four ring potentiometric probe with platinum sensors to offer versatility over typical amperometric designs. By utilizing the four ring method. it is possible to measure very low or high conductivity levels without changing probes. Three options of compensating for temperature are available for this instrument:        \u003c\/p\u003e\n\u003cp style=\"text-align: justify;\"\u003eAutomatic (ATC): The EC probe has a built-in temperature sensor which is used to automatically compensate the EC\/TDS reading (from -9.9°C to 120.0°C). using the selected reference temperature (20 or 25ºC) and temperature compensation coefficient from (0.0 to 6.0%)\/ºC\u003c\/p\u003e\n\u003cp style=\"text-align: justify;\"\u003eManual (MTC): The temperature value. shown on the secondary LCD. can be manually set with the ARROW keys. The compensation is referenced at the selected temperature. All the other parameters of temperature compensation are settable similar to ATC.\u003c\/p\u003e\n\u003cp style=\"text-align: justify;\"\u003eNo Compensation (NoTC): For actual conductivity or TDS measurement. the temperature value shown on the secondary LCD is not taken into account.\u003c\/p\u003e\n\u003cp style=\"text-align: justify;\"\u003eCell constant is selectable between 0.5 and 1.700. TDS factor is selectable between 0.40 and 0.80.\u003c\/p\u003e\n\u003cp style=\"text-align: justify;\"\u003eThe HI 2300 also provides users with GLP capabilities. Good Laboratory Practice (GLP) is a set of functions that allows storage and retrieval of data regarding the status of the system.\u003c\/p\u003e\n\u003cp style=\"text-align: justify;\"\u003eAfter a successful calibration. the meter automatically stores the date and time of calibration. the calibration solution used and the resulting cell constant value. All this information can be later recalled by the user. Other features include a lock range function and stability indicator.\u003c\/p\u003e\n\u003cp style=\"text-align: justify;\"\u003eFor PC communication. use the optional HI 92000 software and HI 920013 USB cable. The software is provided with an exclusive online guide of all the commands available and allows data printing. plotting and exporting.\u003c\/p\u003e\n\u003cp style=\"text-align: justify;\"\u003e \u003c\/p\u003e\n\u003cp style=\"text-align: justify;\"\u003e\u003cstrong\u003eOrder Information:\u003c\/strong\u003e   HI 2300-02 (230V) is supplied with HI 76310 conductivity probe. 12 VDC adapter and instructions.\u003c\/p\u003e\n\u003cdiv\u003e\u003cstrong\u003eProduct Manuals\u003c\/strong\u003e\u003c\/div\u003e\n\u003cp\u003e\u003cimg width=\"20\" height=\"22\" align=\"absmiddle\" alt=\"\" src=\"http:\/\/www.hannainst.com\/usa\/images\/icon-pdf.gif\" border=\"0\"\u003e  Manual: \u003ca title=\"Download Product Manual\" href=\"http:\/\/www.hannainst.com\/manuals\/man2300Y_09_12.pdf\" target=\"_blank\" rel=\"noopener noreferrer\"\u003eDownload\u003c\/a\u003e\u003c\/p\u003e\n\u003cp\u003e\u003cstrong\u003eSpecification :\u003c\/strong\u003e\u003c\/p\u003e\n\u003ctable class=\"spec_table\"\u003e\n\u003ctbody\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eRange\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e\u003cstrong\u003e\u003csmall\u003eEC\u003c\/small\u003e\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e0.00 to 29.99 µS\/cm; 30.0 to 299.9 µS\/cm; 300 to 2999 µS\/cm; 3.00 to 29.99 mS\/cm; 30.0 to 200.0 mS\/cm; up to 500.0 mS\/cm (actual EC)\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003cstrong\u003e\u003csmall\u003eTDS\u003c\/small\u003e\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e0.00 to 14.99 mg\/L (ppm); 15.0 to 149.9 mg\/L (ppm); 150 to 1499 mg\/L (ppm); 1.50 to 14.99 g\/L (ppt); 15.0 to 100.0 g\/L (ppt); up to 400.0 g\/L (actual TDS). with 0.80 conversion factor\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003cstrong\u003e\u003csmall\u003eNaCl\u003c\/small\u003e\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e0.0 to 400.0%\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003cstrong\u003e\u003csmall\u003eTemperature\u003c\/small\u003e\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e-20.0 to 120.0°C\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eResolution\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e\u003cstrong\u003e\u003csmall\u003eEC\u003c\/small\u003e\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e0.01 µS\/cm; 0.1 µS\/cm; 1 µS\/cm; 0.01 mS\/cm; 0.1 mS\/cm \u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003cstrong\u003e\u003csmall\u003eTDS\u003c\/small\u003e\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e0.01 mg\/L; 0.1 mg\/L; 1 mg\/L; 0.01 g\/L; 0.1 g\/L \u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003cstrong\u003e\u003csmall\u003eNaCl\u003c\/small\u003e\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e0.1%\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003cstrong\u003e\u003csmall\u003eTemperature\u003c\/small\u003e\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e0.1°C\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eAccuracy @ 25°C\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e\u003cstrong\u003e\u003csmall\u003eEC\u003c\/small\u003e\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e±1% of reading ± (0.05 µS\/cm or 1 digit)\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003cstrong\u003e\u003csmall\u003eTDS\u003c\/small\u003e\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e±1% of reading ± (0.03 mg\/L or 1 digit)\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003cstrong\u003e\u003csmall\u003eNaCl\u003c\/small\u003e\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e±1% of reading\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003cstrong\u003e\u003csmall\u003eTemperature\u003c\/small\u003e\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e±0.4°C\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eCalibration\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e\u003cstrong\u003e\u003csmall\u003eEC\u003c\/small\u003e\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003eautomatic. one point with six memorized values (84. 1413. 5000. 12880. 80000. 111800 µS\/cm) \u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003cstrong\u003e\u003csmall\u003eNaCl\u003c\/small\u003e\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003eone point. with HI 7037 calibration solution\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003cstrong\u003e\u003csmall\u003eTemperature\u003c\/small\u003e\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003etwo point. at 0 and 50°C\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eTemperature Compensation\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003eautomatic or manual from -20.0 to 120.0°C\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eTemperature Coefficient\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003eselectable from 0.00 to 6.00%\/°C (EC and TDS only)\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eTDS Factor\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003eselectable from 0.40 to 0.80 (default value: 0.50)\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eProbe\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003eHI 76310 platinum. four ring conductivity\/TDS probe with internal temperature sensor and 1 m (3.3') cable (included)\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003ePC Connection\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003eopto-isolated USB\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eLogging\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003elog on demand. 500 samples\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eAuto-off\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003eafter five minutes of non-use (can be disabled)\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003ePower Supply\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e12 VDC adapter (included)\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eEnvironment\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e0 to 50°C (32 to 122°F); RH max 95%\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eDimensions\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e235 x 218 x 108 mm (9.2 x 8.5 x 4.2”)\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eWeight\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e 1.3 kg (2.9 lbs.)\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003c\/tbody\u003e\n\u003c\/table\u003e\n\u003cp\u003e\u003cstrong\u003e \u003c\/strong\u003e\u003c\/p\u003e"}

HANNA HI2300-02 EC/TDS/NaCl/°C Meter - Autorange

Product Description

HI 2300 measures EC. TDS. NaCl and temperature. In conductivity and TDS ranges (up to 500 mS/cm and 400 g/L respectively) the instrument automatically chooses the best scale to maintain the highest accuracy.

EC calibration is a one-point procedure. Selectable calibration points are 0.00 µS. 84.0 µS. 1413 µS. 5.00 mS. 12.88 mS. 80.0 mS. and 111.8 mS selected according with the expected measurement range. NaCl calibration is a one-point procedure at 100.0% NaCl. Use HI 7037L calibration solution as a 100% NaCl standard solution. This instrument utilizes a four ring potentiometric probe with platinum sensors to offer versatility over typical amperometric designs. By utilizing the four ring method. it is possible to measure very low or high conductivity levels without changing probes. Three options of compensating for temperature are available for this instrument:        

Automatic (ATC): The EC probe has a built-in temperature sensor which is used to automatically compensate the EC/TDS reading (from -9.9°C to 120.0°C). using the selected reference temperature (20 or 25ºC) and temperature compensation coefficient from (0.0 to 6.0%)/ºC

Manual (MTC): The temperature value. shown on the secondary LCD. can be manually set with the ARROW keys. The compensation is referenced at the selected temperature. All the other parameters of temperature compensation are settable similar to ATC.

No Compensation (NoTC): For actual conductivity or TDS measurement. the temperature value shown on the secondary LCD is not taken into account.

Cell constant is selectable between 0.5 and 1.700. TDS factor is selectable between 0.40 and 0.80.

The HI 2300 also provides users with GLP capabilities. Good Laboratory Practice (GLP) is a set of functions that allows storage and retrieval of data regarding the status of the system.

After a successful calibration. the meter automatically stores the date and time of calibration. the calibration solution used and the resulting cell constant value. All this information can be later recalled by the user. Other features include a lock range function and stability indicator.

For PC communication. use the optional HI 92000 software and HI 920013 USB cable. The software is provided with an exclusive online guide of all the commands available and allows data printing. plotting and exporting.

 

Order Information:   HI 2300-02 (230V) is supplied with HI 76310 conductivity probe. 12 VDC adapter and instructions.

Product Manuals

  Manual: Download

Specification :

Range EC 0.00 to 29.99 µS/cm; 30.0 to 299.9 µS/cm; 300 to 2999 µS/cm; 3.00 to 29.99 mS/cm; 30.0 to 200.0 mS/cm; up to 500.0 mS/cm (actual EC)
  TDS 0.00 to 14.99 mg/L (ppm); 15.0 to 149.9 mg/L (ppm); 150 to 1499 mg/L (ppm); 1.50 to 14.99 g/L (ppt); 15.0 to 100.0 g/L (ppt); up to 400.0 g/L (actual TDS). with 0.80 conversion factor
  NaCl 0.0 to 400.0%
  Temperature -20.0 to 120.0°C
Resolution EC 0.01 µS/cm; 0.1 µS/cm; 1 µS/cm; 0.01 mS/cm; 0.1 mS/cm
  TDS 0.01 mg/L; 0.1 mg/L; 1 mg/L; 0.01 g/L; 0.1 g/L
  NaCl 0.1%
  Temperature 0.1°C
Accuracy @ 25°C EC ±1% of reading ± (0.05 µS/cm or 1 digit)
  TDS ±1% of reading ± (0.03 mg/L or 1 digit)
  NaCl ±1% of reading
  Temperature ±0.4°C
Calibration EC automatic. one point with six memorized values (84. 1413. 5000. 12880. 80000. 111800 µS/cm)
  NaCl one point. with HI 7037 calibration solution
  Temperature two point. at 0 and 50°C
Temperature Compensation   automatic or manual from -20.0 to 120.0°C
Temperature Coefficient   selectable from 0.00 to 6.00%/°C (EC and TDS only)
TDS Factor   selectable from 0.40 to 0.80 (default value: 0.50)
Probe   HI 76310 platinum. four ring conductivity/TDS probe with internal temperature sensor and 1 m (3.3') cable (included)
PC Connection   opto-isolated USB
Logging   log on demand. 500 samples
Auto-off   after five minutes of non-use (can be disabled)
Power Supply   12 VDC adapter (included)
Environment   0 to 50°C (32 to 122°F); RH max 95%
Dimensions   235 x 218 x 108 mm (9.2 x 8.5 x 4.2”)
Weight   1.3 kg (2.9 lbs.)

 