{"id":2012197701,"title":"HANNA HI 2550-02 pH\/mV\/EC\/TDS\/NaCl\/°C Meter - 2 Channel","handle":"hi2550-02","description":"\u003cp style=\"text-align: justify;\"\u003eHI 2550 is a 2 channel instrument that measures up to 7 parameters. With this single laboratory bench meter you can measure pH. ORP. ISE. conductivity (EC). TDS. NaCl percentage and temperature.\u003c\/p\u003e\n\u003cp style=\"text-align: justify;\"\u003eUtilizing an external temperature probe. pH readings are automatically temperature compensated. To ensure a higher level of precision. pH calibrations are up to five calibration points. chosen from the seven available memorized buffers.\u003c\/p\u003e\n\u003cp style=\"text-align: justify;\"\u003eThis instrument can measure using ORP electrodes (pH channel input). due to it’s capability to measure mV with a resolution up to 0.1 mV and also use ISE electrodes on the mV scale (pH channel input).\u003c\/p\u003e\n\u003cp style=\"text-align: justify;\"\u003eEC measurements can be compensated relative to a selected reference temperature. The EC calibration mode allows you to chose from among six recognized conductivity standards and perform a single-point calibration. The most suitable EC and TDS range for your application is automatically selected. The HI 2550 also includes the ability to set and lock the range manually.\u003c\/p\u003e\n\u003cp style=\"text-align: justify;\"\u003eThis instrument provides GLP capabilities that allows for the storage and retrieval of all data regarding pH. rel mV. EC and NaCl calibration and sample measurement as well as data regarding the maintenance and status of the electrode.\u003c\/p\u003e\n\u003cp style=\"text-align: justify;\"\u003eWith a built-in logging function. measurements are stored in non volatile memory. and can be transferred to a PC through the USB port. Users can manually log up to 200 records and interval log up to 500 records.\u003c\/p\u003e\n\u003cp style=\"text-align: justify;\"\u003e \u003c\/p\u003e\n\u003cp style=\"text-align: justify;\"\u003e\u003cstrong\u003eOrder Information:\u003c\/strong\u003e   HI 2550-02 (230V) are supplied with HI 1131B pH electrode. HI 76310 conductivity\/TDS probe. HI 7662 temperature probe. HI 76404N electrode holder. HI 70004 pH 4.01 buffer solution sachet. HI 70007 pH 7.01 buffer solution sachet. HI 7071S electrolyte solution (30 mL). 12 VDC adapter and instruction manual.\u003c\/p\u003e\n\u003cdiv\u003e\u003cstrong\u003eProduct Manuals\u003c\/strong\u003e\u003c\/div\u003e\n\u003cp style=\"text-align: justify;\"\u003e\u003cimg src=\"http:\/\/www.hannainst.com\/usa\/images\/icon-pdf.gif\" alt=\"\" width=\"20\" height=\"22\" align=\"absmiddle\" border=\"0\"\u003e  Manual: \u003ca title=\"Download Product Manual\" href=\"http:\/\/www.hannainst.com\/manuals\/man2550_23_04_2009.pdf\" target=\"_blank\"\u003eDownload\u003c\/a\u003e\u003c\/p\u003e\n\u003cp\u003e\u003cstrong\u003eSpecification :\u003c\/strong\u003e\u003c\/p\u003e\n\u003ctable class=\"spec_table\"\u003e\n\u003ctbody\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eRange\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e\u003cstrong\u003e\u003csmall\u003epH\u003c\/small\u003e\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e-2.0 to 16.0 pH; -2.00 to 16.00 pH; -2.000 to 16.000 pH\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003cstrong\u003e\u003csmall\u003eISE \u0026amp; ORP\u003c\/small\u003e\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e±999.9 mV (ISE \u0026amp; ORP); ±2000 mV (ISE \u0026amp; ORP)\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003cstrong\u003e\u003csmall\u003eEC\u003c\/small\u003e\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e0.00 to 29.99 µS\/cm; 30.0 to 299.9 µS\/cm; 300 to 2999 µS\/cm;  3.00 to 29.99 mS\/cm; 30.0 to 200.0 mS\/cm; up to 500.0 mS\/cm actual conductivity\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003cstrong\u003e\u003csmall\u003eTDS\u003c\/small\u003e\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e0.00 to 14.99 ppm; 15.0 to 149.9 ppm; 150 to 1499 ppm; 1.50 to 14.99 g\/L; 15.0 to 100.0 g\/L; up to 400.0 g\/L actual TDS (with 0.80 factor)\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003cstrong\u003e\u003csmall\u003eNaCl\u003c\/small\u003e\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e0.0 to 400.0% NaCl\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003cstrong\u003e\u003csmall\u003eTemperature\u003c\/small\u003e\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e–20.0 to 120.0 ºC (pH. EC range)\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eResolution\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e\u003cstrong\u003e\u003csmall\u003epH\u003c\/small\u003e\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e0.1 pH; 0.01 pH; 0.001 pH\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003cstrong\u003e\u003csmall\u003eISE \u0026amp; ORP\u003c\/small\u003e\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e0.1 mV (±999.9 mV); 1 mV (± 2000 mV)\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003cstrong\u003e\u003csmall\u003eTDS\u003c\/small\u003e\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e0.01 ppm; 0.1 ppm; 1 ppm; 0.01 g\/L; 0.1 g\/L\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003cstrong\u003e\u003csmall\u003eNaCl\u003c\/small\u003e\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e0.1% NaCl\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003cstrong\u003e\u003csmall\u003eTemperature\u003c\/small\u003e\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e0.1 ºC\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eAccuracy @ 20°C\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e\u003cstrong\u003e\u003csmall\u003epH\u003c\/small\u003e\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e±0.01 pH; ±0.002 pH\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003cstrong\u003e\u003csmall\u003eISE \u0026amp; ORP\u003c\/small\u003e\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e±0.2 mV (±999.9 mV); ±1 mV (±2000 mV)\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003cstrong\u003e\u003csmall\u003eEC\u003c\/small\u003e\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e±1 % reading (±0.05 µS\/cm or 1 digit. whichever greater)\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003cstrong\u003e\u003csmall\u003eTDS\u003c\/small\u003e\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e±1% of reading (±0.03 ppm or 1 digit. whichever greater)\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003cstrong\u003e\u003csmall\u003eNaCl\u003c\/small\u003e\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e±1% of reading\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003cstrong\u003e\u003csmall\u003eTemperature\u003c\/small\u003e\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e±0.4 ºC (excluding probe error)\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eCalibration\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e\u003cstrong\u003e\u003csmall\u003epH\u003c\/small\u003e\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003eup to five point calibration. seven standard buffers available (1.68. 4.01. 6.86. 7.01. 9.18. 10.01. 12.45). and two custom buffers\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003cstrong\u003e\u003csmall\u003eEC\u003c\/small\u003e\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003etwo point calibration; one point slope calibration; six buffers available: 84.0. 1413 µS\/cm; 5.00. 12.88. 80.0. 111.8 mS\/cm; one point offset: 0.00 µS\/cm\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003cstrong\u003e\u003csmall\u003eNaCl\u003c\/small\u003e\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003eone point with HI 7037L standard (optional)\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eProbes\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e\u003cstrong\u003e\u003csmall\u003epH\u003c\/small\u003e\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003eHI 1131B glass body pH electrode with BNC connector and 1 m (3.3') cable (included)\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003cstrong\u003e\u003csmall\u003eEC\u003c\/small\u003e\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003eHI 76310 platinum four ring conductivity\/TDS probe with built-in temperature sensor and 1 m (3.3') cable (included)\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003cstrong\u003e\u003csmall\u003eTemperature\u003c\/small\u003e\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003eHI 7662 temperature probe with 1 m (3.3') cable (included)\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eRelative mV Offset\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e±2000 mV\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eTemperature Compensation\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003emanual or automatic from: -20.0 to 120.0 ºC (pH range) -20.0 to 120.0 ºC (EC range) (can be disabled on conductivity range to measure actual conductivity)\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eCond. Temp. Coefficient\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e0.00 to 6.00 %\/ºC (for EC and TDS only) default value is 1.90 %\/ºC\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eTDS Factor\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e0.40 to 0.80 (default value is 0.50)\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eInput Impedance\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e10\u003csup\u003e12\u003c\/sup\u003e ohms\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003ePC Connectivity\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003eopto-isolated USB\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eLog on Demand\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e200 records\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eLog Interval Feature\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e500 records; 5. 10. 30 sec and 1. 2. 5. 10. 15. 30. 60. 120. 180 min stability logging (“StAb”)\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003ePower Supply\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e12 VDC\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eEnvironment\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e0 to 50°C (32 to 122°F); RH max 95% non-condensing\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eDimensions\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e235 x 218 x 108 mm (9.2 x 8.5 x 4.2”)\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eWeight\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e1.3 Kg (2.9 lb); kit with holder 2.1 Kg (4.6 lb.)\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003c\/tbody\u003e\n\u003c\/table\u003e\n\u003cp\u003e\u003cstrong\u003e \u003c\/strong\u003e\u003c\/p\u003e","published_at":"2026-06-25T15:21:40+12:00","created_at":"2015-09-04T15:23:18+12:00","vendor":"Hanna Instruments","type":"Electrochemical","tags":["Manufacturer_Hanna Instruments","Type_Conductivity\/TDS Meters","Type_pH \u0026 ORP Meters"],"price":186200,"price_min":186200,"price_max":186200,"available":true,"price_varies":false,"compare_at_price":null,"compare_at_price_min":0,"compare_at_price_max":0,"compare_at_price_varies":false,"variants":[{"id":5764491909,"title":"Default Title","option1":"Default Title","option2":null,"option3":null,"sku":"HI2550-02","requires_shipping":false,"taxable":true,"featured_image":null,"available":true,"name":"HANNA HI 2550-02 pH\/mV\/EC\/TDS\/NaCl\/°C Meter - 2 Channel","public_title":null,"options":["Default Title"],"price":186200,"weight":0,"compare_at_price":null,"inventory_quantity":1,"inventory_management":null,"inventory_policy":"deny","barcode":null,"requires_selling_plan":false,"selling_plan_allocations":[],"quantity_rule":{"min":1,"max":null,"increment":1}}],"images":["\/\/acornsci.co.nz\/cdn\/shop\/products\/ph-mv-ec-tds-nacl-c-metre-fonctions-bpl-236_1-1_2015.jpeg?v=1465536564","\/\/acornsci.co.nz\/cdn\/shop\/products\/hi2550-1_2015.jpeg?v=1465536565"],"featured_image":"\/\/acornsci.co.nz\/cdn\/shop\/products\/ph-mv-ec-tds-nacl-c-metre-fonctions-bpl-236_1-1_2015.jpeg?v=1465536564","options":["Title"],"media":[{"alt":"HI 2550-02 pH\/mV\/EC\/TDS\/NaCl\/°C Meter - 2 Channel - Acorn Scientific","id":25187123299,"position":1,"preview_image":{"aspect_ratio":1.0,"height":1000,"width":1000,"src":"\/\/acornsci.co.nz\/cdn\/shop\/products\/ph-mv-ec-tds-nacl-c-metre-fonctions-bpl-236_1-1_2015.jpeg?v=1465536564"},"aspect_ratio":1.0,"height":1000,"media_type":"image","src":"\/\/acornsci.co.nz\/cdn\/shop\/products\/ph-mv-ec-tds-nacl-c-metre-fonctions-bpl-236_1-1_2015.jpeg?v=1465536564","width":1000},{"alt":"HI 2550-02 pH\/mV\/EC\/TDS\/NaCl\/°C Meter - 2 Channel - Acorn Scientific","id":25187156067,"position":2,"preview_image":{"aspect_ratio":1.0,"height":800,"width":800,"src":"\/\/acornsci.co.nz\/cdn\/shop\/products\/hi2550-1_2015.jpeg?v=1465536565"},"aspect_ratio":1.0,"height":800,"media_type":"image","src":"\/\/acornsci.co.nz\/cdn\/shop\/products\/hi2550-1_2015.jpeg?v=1465536565","width":800}],"requires_selling_plan":false,"selling_plan_groups":[],"content":"\u003cp style=\"text-align: justify;\"\u003eHI 2550 is a 2 channel instrument that measures up to 7 parameters. With this single laboratory bench meter you can measure pH. ORP. ISE. conductivity (EC). TDS. NaCl percentage and temperature.\u003c\/p\u003e\n\u003cp style=\"text-align: justify;\"\u003eUtilizing an external temperature probe. pH readings are automatically temperature compensated. To ensure a higher level of precision. pH calibrations are up to five calibration points. chosen from the seven available memorized buffers.\u003c\/p\u003e\n\u003cp style=\"text-align: justify;\"\u003eThis instrument can measure using ORP electrodes (pH channel input). due to it’s capability to measure mV with a resolution up to 0.1 mV and also use ISE electrodes on the mV scale (pH channel input).\u003c\/p\u003e\n\u003cp style=\"text-align: justify;\"\u003eEC measurements can be compensated relative to a selected reference temperature. The EC calibration mode allows you to chose from among six recognized conductivity standards and perform a single-point calibration. The most suitable EC and TDS range for your application is automatically selected. The HI 2550 also includes the ability to set and lock the range manually.\u003c\/p\u003e\n\u003cp style=\"text-align: justify;\"\u003eThis instrument provides GLP capabilities that allows for the storage and retrieval of all data regarding pH. rel mV. EC and NaCl calibration and sample measurement as well as data regarding the maintenance and status of the electrode.\u003c\/p\u003e\n\u003cp style=\"text-align: justify;\"\u003eWith a built-in logging function. measurements are stored in non volatile memory. and can be transferred to a PC through the USB port. Users can manually log up to 200 records and interval log up to 500 records.\u003c\/p\u003e\n\u003cp style=\"text-align: justify;\"\u003e \u003c\/p\u003e\n\u003cp style=\"text-align: justify;\"\u003e\u003cstrong\u003eOrder Information:\u003c\/strong\u003e   HI 2550-02 (230V) are supplied with HI 1131B pH electrode. HI 76310 conductivity\/TDS probe. HI 7662 temperature probe. HI 76404N electrode holder. HI 70004 pH 4.01 buffer solution sachet. HI 70007 pH 7.01 buffer solution sachet. HI 7071S electrolyte solution (30 mL). 12 VDC adapter and instruction manual.\u003c\/p\u003e\n\u003cdiv\u003e\u003cstrong\u003eProduct Manuals\u003c\/strong\u003e\u003c\/div\u003e\n\u003cp style=\"text-align: justify;\"\u003e\u003cimg src=\"http:\/\/www.hannainst.com\/usa\/images\/icon-pdf.gif\" alt=\"\" width=\"20\" height=\"22\" align=\"absmiddle\" border=\"0\"\u003e  Manual: \u003ca title=\"Download Product Manual\" href=\"http:\/\/www.hannainst.com\/manuals\/man2550_23_04_2009.pdf\" target=\"_blank\"\u003eDownload\u003c\/a\u003e\u003c\/p\u003e\n\u003cp\u003e\u003cstrong\u003eSpecification :\u003c\/strong\u003e\u003c\/p\u003e\n\u003ctable class=\"spec_table\"\u003e\n\u003ctbody\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eRange\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e\u003cstrong\u003e\u003csmall\u003epH\u003c\/small\u003e\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e-2.0 to 16.0 pH; -2.00 to 16.00 pH; -2.000 to 16.000 pH\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003cstrong\u003e\u003csmall\u003eISE \u0026amp; ORP\u003c\/small\u003e\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e±999.9 mV (ISE \u0026amp; ORP); ±2000 mV (ISE \u0026amp; ORP)\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003cstrong\u003e\u003csmall\u003eEC\u003c\/small\u003e\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e0.00 to 29.99 µS\/cm; 30.0 to 299.9 µS\/cm; 300 to 2999 µS\/cm;  3.00 to 29.99 mS\/cm; 30.0 to 200.0 mS\/cm; up to 500.0 mS\/cm actual conductivity\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003cstrong\u003e\u003csmall\u003eTDS\u003c\/small\u003e\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e0.00 to 14.99 ppm; 15.0 to 149.9 ppm; 150 to 1499 ppm; 1.50 to 14.99 g\/L; 15.0 to 100.0 g\/L; up to 400.0 g\/L actual TDS (with 0.80 factor)\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003cstrong\u003e\u003csmall\u003eNaCl\u003c\/small\u003e\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e0.0 to 400.0% NaCl\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003cstrong\u003e\u003csmall\u003eTemperature\u003c\/small\u003e\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e–20.0 to 120.0 ºC (pH. EC range)\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eResolution\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e\u003cstrong\u003e\u003csmall\u003epH\u003c\/small\u003e\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e0.1 pH; 0.01 pH; 0.001 pH\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003cstrong\u003e\u003csmall\u003eISE \u0026amp; ORP\u003c\/small\u003e\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e0.1 mV (±999.9 mV); 1 mV (± 2000 mV)\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003cstrong\u003e\u003csmall\u003eTDS\u003c\/small\u003e\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e0.01 ppm; 0.1 ppm; 1 ppm; 0.01 g\/L; 0.1 g\/L\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003cstrong\u003e\u003csmall\u003eNaCl\u003c\/small\u003e\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e0.1% NaCl\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003cstrong\u003e\u003csmall\u003eTemperature\u003c\/small\u003e\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e0.1 ºC\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eAccuracy @ 20°C\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e\u003cstrong\u003e\u003csmall\u003epH\u003c\/small\u003e\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e±0.01 pH; ±0.002 pH\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003cstrong\u003e\u003csmall\u003eISE \u0026amp; ORP\u003c\/small\u003e\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e±0.2 mV (±999.9 mV); ±1 mV (±2000 mV)\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003cstrong\u003e\u003csmall\u003eEC\u003c\/small\u003e\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e±1 % reading (±0.05 µS\/cm or 1 digit. whichever greater)\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003cstrong\u003e\u003csmall\u003eTDS\u003c\/small\u003e\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e±1% of reading (±0.03 ppm or 1 digit. whichever greater)\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003cstrong\u003e\u003csmall\u003eNaCl\u003c\/small\u003e\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e±1% of reading\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003cstrong\u003e\u003csmall\u003eTemperature\u003c\/small\u003e\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e±0.4 ºC (excluding probe error)\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eCalibration\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e\u003cstrong\u003e\u003csmall\u003epH\u003c\/small\u003e\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003eup to five point calibration. seven standard buffers available (1.68. 4.01. 6.86. 7.01. 9.18. 10.01. 12.45). and two custom buffers\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003cstrong\u003e\u003csmall\u003eEC\u003c\/small\u003e\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003etwo point calibration; one point slope calibration; six buffers available: 84.0. 1413 µS\/cm; 5.00. 12.88. 80.0. 111.8 mS\/cm; one point offset: 0.00 µS\/cm\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003cstrong\u003e\u003csmall\u003eNaCl\u003c\/small\u003e\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003eone point with HI 7037L standard (optional)\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eProbes\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e\u003cstrong\u003e\u003csmall\u003epH\u003c\/small\u003e\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003eHI 1131B glass body pH electrode with BNC connector and 1 m (3.3') cable (included)\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003cstrong\u003e\u003csmall\u003eEC\u003c\/small\u003e\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003eHI 76310 platinum four ring conductivity\/TDS probe with built-in temperature sensor and 1 m (3.3') cable (included)\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003cstrong\u003e\u003csmall\u003eTemperature\u003c\/small\u003e\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003eHI 7662 temperature probe with 1 m (3.3') cable (included)\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eRelative mV Offset\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e±2000 mV\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eTemperature Compensation\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003emanual or automatic from: -20.0 to 120.0 ºC (pH range) -20.0 to 120.0 ºC (EC range) (can be disabled on conductivity range to measure actual conductivity)\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eCond. Temp. Coefficient\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e0.00 to 6.00 %\/ºC (for EC and TDS only) default value is 1.90 %\/ºC\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eTDS Factor\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e0.40 to 0.80 (default value is 0.50)\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eInput Impedance\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e10\u003csup\u003e12\u003c\/sup\u003e ohms\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003ePC Connectivity\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003eopto-isolated USB\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eLog on Demand\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e200 records\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eLog Interval Feature\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e500 records; 5. 10. 30 sec and 1. 2. 5. 10. 15. 30. 60. 120. 180 min stability logging (“StAb”)\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003ePower Supply\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e12 VDC\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eEnvironment\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e0 to 50°C (32 to 122°F); RH max 95% non-condensing\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eDimensions\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e235 x 218 x 108 mm (9.2 x 8.5 x 4.2”)\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eWeight\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e1.3 Kg (2.9 lb); kit with holder 2.1 Kg (4.6 lb.)\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003c\/tbody\u003e\n\u003c\/table\u003e\n\u003cp\u003e\u003cstrong\u003e \u003c\/strong\u003e\u003c\/p\u003e"}

HANNA HI 2550-02 pH/mV/EC/TDS/NaCl/°C Meter - 2 Channel

Product Description

HI 2550 is a 2 channel instrument that measures up to 7 parameters. With this single laboratory bench meter you can measure pH. ORP. ISE. conductivity (EC). TDS. NaCl percentage and temperature.

Utilizing an external temperature probe. pH readings are automatically temperature compensated. To ensure a higher level of precision. pH calibrations are up to five calibration points. chosen from the seven available memorized buffers.

This instrument can measure using ORP electrodes (pH channel input). due to it’s capability to measure mV with a resolution up to 0.1 mV and also use ISE electrodes on the mV scale (pH channel input).

EC measurements can be compensated relative to a selected reference temperature. The EC calibration mode allows you to chose from among six recognized conductivity standards and perform a single-point calibration. The most suitable EC and TDS range for your application is automatically selected. The HI 2550 also includes the ability to set and lock the range manually.

This instrument provides GLP capabilities that allows for the storage and retrieval of all data regarding pH. rel mV. EC and NaCl calibration and sample measurement as well as data regarding the maintenance and status of the electrode.

With a built-in logging function. measurements are stored in non volatile memory. and can be transferred to a PC through the USB port. Users can manually log up to 200 records and interval log up to 500 records.

 

Order Information:   HI 2550-02 (230V) are supplied with HI 1131B pH electrode. HI 76310 conductivity/TDS probe. HI 7662 temperature probe. HI 76404N electrode holder. HI 70004 pH 4.01 buffer solution sachet. HI 70007 pH 7.01 buffer solution sachet. HI 7071S electrolyte solution (30 mL). 12 VDC adapter and instruction manual.

Product Manuals

  Manual: Download

Specification :

Range pH -2.0 to 16.0 pH; -2.00 to 16.00 pH; -2.000 to 16.000 pH
  ISE & ORP ±999.9 mV (ISE & ORP); ±2000 mV (ISE & ORP)
  EC 0.00 to 29.99 µS/cm; 30.0 to 299.9 µS/cm; 300 to 2999 µS/cm;  3.00 to 29.99 mS/cm; 30.0 to 200.0 mS/cm; up to 500.0 mS/cm actual conductivity
  TDS 0.00 to 14.99 ppm; 15.0 to 149.9 ppm; 150 to 1499 ppm; 1.50 to 14.99 g/L; 15.0 to 100.0 g/L; up to 400.0 g/L actual TDS (with 0.80 factor)
  NaCl 0.0 to 400.0% NaCl
  Temperature –20.0 to 120.0 ºC (pH. EC range)
Resolution pH 0.1 pH; 0.01 pH; 0.001 pH
  ISE & ORP 0.1 mV (±999.9 mV); 1 mV (± 2000 mV)
  TDS 0.01 ppm; 0.1 ppm; 1 ppm; 0.01 g/L; 0.1 g/L
  NaCl 0.1% NaCl
  Temperature 0.1 ºC
Accuracy @ 20°C pH ±0.01 pH; ±0.002 pH
  ISE & ORP ±0.2 mV (±999.9 mV); ±1 mV (±2000 mV)
  EC ±1 % reading (±0.05 µS/cm or 1 digit. whichever greater)
  TDS ±1% of reading (±0.03 ppm or 1 digit. whichever greater)
  NaCl ±1% of reading
  Temperature ±0.4 ºC (excluding probe error)
Calibration pH up to five point calibration. seven standard buffers available (1.68. 4.01. 6.86. 7.01. 9.18. 10.01. 12.45). and two custom buffers
  EC two point calibration; one point slope calibration; six buffers available: 84.0. 1413 µS/cm; 5.00. 12.88. 80.0. 111.8 mS/cm; one point offset: 0.00 µS/cm
  NaCl one point with HI 7037L standard (optional)
Probes pH HI 1131B glass body pH electrode with BNC connector and 1 m (3.3') cable (included)
  EC HI 76310 platinum four ring conductivity/TDS probe with built-in temperature sensor and 1 m (3.3') cable (included)
  Temperature HI 7662 temperature probe with 1 m (3.3') cable (included)
Relative mV Offset   ±2000 mV
Temperature Compensation   manual or automatic from: -20.0 to 120.0 ºC (pH range) -20.0 to 120.0 ºC (EC range) (can be disabled on conductivity range to measure actual conductivity)
Cond. Temp. Coefficient   0.00 to 6.00 %/ºC (for EC and TDS only) default value is 1.90 %/ºC
TDS Factor   0.40 to 0.80 (default value is 0.50)
Input Impedance   1012 ohms
PC Connectivity   opto-isolated USB
Log on Demand   200 records
Log Interval Feature   500 records; 5. 10. 30 sec and 1. 2. 5. 10. 15. 30. 60. 120. 180 min stability logging (“StAb”)
Power Supply   12 VDC
Environment   0 to 50°C (32 to 122°F); RH max 95% non-condensing
Dimensions   235 x 218 x 108 mm (9.2 x 8.5 x 4.2”)
Weight   1.3 Kg (2.9 lb); kit with holder 2.1 Kg (4.6 lb.)

 

This product is available upon request. Contact us for more information.