{"id":1582122309,"title":"HANNA HI3222-02 pH \/ ORP \/ ISE \/ °C Meter \/ CalCheck™ + PC Interface - 2 Channel","handle":"hi3222-02","description":"\u003cp style=\"text-align: justify;\"\u003eHANNA’s HI 3222 benchtop instrument features up to five point pH calibration with a choice of five custom buffers and seven standard buffers.\u003c\/p\u003e\n\u003cp style=\"text-align: justify;\"\u003eHI 3222 features an interactive user support interface that assists you before. during and after measurement. On-screen tutorials guide users through set-up. calibration and measurement while context sensitive help of any screen is available at a push of a button. The help screen includes language specific assistance for menu parameters. calibration. logging. contact and accessory information for your instrument.\u003c\/p\u003e\n\u003cp style=\"text-align: justify;\"\u003eThis instrument features HANNA’s exclusive Calibration Check™. a diagnostics system that ensures accurate pH readings every time. By alerting users of potential problems during the calibration process. the Calibration Check™ system eliminates erroneous readings due to dirty or faulty pH electrodes or contaminated pH buffer solutions during calibration. Throughout the calibration process. users are guided step-by-step by the on-screen tutorial. After calibration. the probe condition is evaluated and an indicator is displayed informing the user of the overall pH electrode status.\u003c\/p\u003e\n\u003cp style=\"text-align: justify;\"\u003eHI 3222 is equipped with two input channels for simultaneous measurements. Having these two channels eliminates the need for swapping probes and recalibrating.\u003c\/p\u003e\n\u003cp style=\"text-align: justify;\"\u003eThis meter can measure using ORP electrodes (pH channel input). thanks to it's capability to measure mV with a resolution up to 0.1 mV. The HI 3222 can use ISE electrodes in the ppm scale (pH channel input) and provides a choice of measurement units (ppb. ppm. molarity. weight\/volume %). The electrode type and unit selection capability and the ISE calibration in up to five calibration standard solutions make these instruments very useful for a large range of concentration solutions measurements.\u003c\/p\u003e\n\u003cp style=\"text-align: justify;\"\u003e \u003c\/p\u003e\n\u003cp style=\"text-align: justify;\"\u003e\u003cstrong\u003eOrder Information:\u003c\/strong\u003e HI 3222-02 (230V) are supplied with HI 1131B pH electrode. HI 7662-T temperature probe. HI 76404N electrode holder. HI 70004 pH 4.01 buffer solution sachet. HI 70007 pH 7.01 buffer solution sachet. HI 700661 electrode cleaning solution sachet (2). HI 7071S electrolyte solution (30 mL). 12 VDC adapter and instructions.\u003c\/p\u003e\n\u003cdiv\u003e\u003cstrong\u003eProduct Manuals\u003c\/strong\u003e\u003c\/div\u003e\n\u003cp style=\"text-align: justify;\"\u003e\u003cimg src=\"http:\/\/www.hannainst.com\/usa\/images\/icon-pdf.gif\" alt=\"\" width=\"20\" height=\"22\" align=\"absmiddle\" border=\"0\"\u003e Manual: \u003ca title=\"Download Product Manual\" href=\"http:\/\/www.hannainst.com\/manuals\/manHI_3220_3221_3222.pdf\" target=\"_blank\" rel=\"noopener noreferrer\"\u003eDownload\u003c\/a\u003e\u003c\/p\u003e\n\u003cp\u003e\u003cstrong\u003eSpecification :\u003c\/strong\u003e\u003c\/p\u003e\n\u003ctable class=\"spec_table\"\u003e\n\u003ctbody\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eRange\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003epH\u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e-2.0 to 20.0; -2.00 to 20.00; -2.000 to 20.000 pH\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003emV\u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e±2000 mV\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003eISE\u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e1.00 E-7 to 9.99 E10 concentration (choice of units)\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003eTemperature\u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e-20.0 to 120.0 °C (-4.0 to 248.0°F)\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eResolution\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003epH\u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e0.1; 0.01; 0.001 pH\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003emV\u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e0.1 mV\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003eISE\u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e3 digits 0.01; 0.1; 1; 10 concentration\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003eTemperature\u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e0.1°C (0.1°F)\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eAccuracy \u003csmall\u003e(@ 20ºC\/68ºF)\u003c\/small\u003e\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003epH\u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e±0.01; ±0.002 pH\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003emV\u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e±0.2 mV\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003eISE\u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e±0.5% of reading (monovalent ions). ±1% of reading (divalent ions)\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003eTemperature\u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e±0.2°C (±0.4°F) (excluding probe error)\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eCalibration\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003epH\u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003eup to five point calibration. seven standard buffers available (1.68. 4.01. 6.86. 7.01. 9.18. 10.01. 12.45) + five custom buffers\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003eISE\u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003eup to five point calibration. six standard solutions (in units selected)\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003eSlope\u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003efrom 80 to 110%\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003epH Calibration Check™\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003eyes\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eRel. mV Offset Range\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e±2000 mV\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003epH Temp. Compensation\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003emanual or automatic from -20.0 to 120.0°C (-4.0 to 248.0°F)\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eInput Channels\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e2\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003epH Electrode\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003eHI 1131B pH electrode with glass body. BNC connector and 1 m (3.3’) cable (included)\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eTemperature Probe\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003eHI 7662-T temperature probe. stainless steel with 1 m (3.3’) cable (included)\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eLogging\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003elog on demand 400 samples\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eLot Logging\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e5. 10. 30 seconds; 1. 2. 5. 10. 15. 30. 60. 120. 180 minutes (max 600 samples)\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003ePC Connectivity\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003eopto-isolated USB (with optional HI 92000 software)\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eInput Impedance\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e10\u003csup\u003e12\u003c\/sup\u003e Ohms\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003ePower Supply\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e12 VDC adapter (included)\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eEnvironment\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e0-50°C (32 to 122°F) RH max 55% non-condensing\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eDimensions\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e235 x 207 x 110 mm (9.2 x 8.14 x 4.33”)\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eWeight\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e1.8 kg (4.1 lbs.)\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003c\/tbody\u003e\n\u003c\/table\u003e\n\u003cp\u003e\u003cstrong\u003e \u003c\/strong\u003e\u003c\/p\u003e","published_at":"2016-12-14T09:55:00+13:00","created_at":"2015-09-01T15:10:35+12:00","vendor":"Hanna Instruments","type":"Electrochemical","tags":["Category_Electrochemical","Manufacturer_Hanna Instruments","Type_pH \u0026 ORP Meters"],"price":0,"price_min":0,"price_max":0,"available":true,"price_varies":false,"compare_at_price":0,"compare_at_price_min":0,"compare_at_price_max":0,"compare_at_price_varies":false,"variants":[{"id":4857214725,"title":"Default Title","option1":"Default Title","option2":null,"option3":null,"sku":"HI3222-02","requires_shipping":false,"taxable":true,"featured_image":null,"available":true,"name":"HANNA HI3222-02 pH \/ ORP \/ ISE \/ °C Meter \/ CalCheck™ + PC Interface - 2 Channel","public_title":null,"options":["Default Title"],"price":0,"weight":0,"compare_at_price":0,"inventory_quantity":1,"inventory_management":null,"inventory_policy":"deny","barcode":null,"requires_selling_plan":false,"selling_plan_allocations":[]}],"images":["\/\/acornsci.co.nz\/cdn\/shop\/products\/labor-phmv-meter-mit-pc-interface-und-cal-check-249_1-1_2015_5630ba65-ff45-4f89-ade7-a68ce522b86e.jpeg?v=1465536610"],"featured_image":"\/\/acornsci.co.nz\/cdn\/shop\/products\/labor-phmv-meter-mit-pc-interface-und-cal-check-249_1-1_2015_5630ba65-ff45-4f89-ade7-a68ce522b86e.jpeg?v=1465536610","options":["Title"],"media":[{"alt":"HI 3222-02 pH \/ ORP \/ ISE \/ °C Meter \/ CalCheck™ + PC Interface - 2 Channel - Acorn Scientific","id":21285994595,"position":1,"preview_image":{"aspect_ratio":1.0,"height":1000,"width":1000,"src":"\/\/acornsci.co.nz\/cdn\/shop\/products\/labor-phmv-meter-mit-pc-interface-und-cal-check-249_1-1_2015_5630ba65-ff45-4f89-ade7-a68ce522b86e.jpeg?v=1465536610"},"aspect_ratio":1.0,"height":1000,"media_type":"image","src":"\/\/acornsci.co.nz\/cdn\/shop\/products\/labor-phmv-meter-mit-pc-interface-und-cal-check-249_1-1_2015_5630ba65-ff45-4f89-ade7-a68ce522b86e.jpeg?v=1465536610","width":1000}],"requires_selling_plan":false,"selling_plan_groups":[],"content":"\u003cp style=\"text-align: justify;\"\u003eHANNA’s HI 3222 benchtop instrument features up to five point pH calibration with a choice of five custom buffers and seven standard buffers.\u003c\/p\u003e\n\u003cp style=\"text-align: justify;\"\u003eHI 3222 features an interactive user support interface that assists you before. during and after measurement. On-screen tutorials guide users through set-up. calibration and measurement while context sensitive help of any screen is available at a push of a button. The help screen includes language specific assistance for menu parameters. calibration. logging. contact and accessory information for your instrument.\u003c\/p\u003e\n\u003cp style=\"text-align: justify;\"\u003eThis instrument features HANNA’s exclusive Calibration Check™. a diagnostics system that ensures accurate pH readings every time. By alerting users of potential problems during the calibration process. the Calibration Check™ system eliminates erroneous readings due to dirty or faulty pH electrodes or contaminated pH buffer solutions during calibration. Throughout the calibration process. users are guided step-by-step by the on-screen tutorial. After calibration. the probe condition is evaluated and an indicator is displayed informing the user of the overall pH electrode status.\u003c\/p\u003e\n\u003cp style=\"text-align: justify;\"\u003eHI 3222 is equipped with two input channels for simultaneous measurements. Having these two channels eliminates the need for swapping probes and recalibrating.\u003c\/p\u003e\n\u003cp style=\"text-align: justify;\"\u003eThis meter can measure using ORP electrodes (pH channel input). thanks to it's capability to measure mV with a resolution up to 0.1 mV. The HI 3222 can use ISE electrodes in the ppm scale (pH channel input) and provides a choice of measurement units (ppb. ppm. molarity. weight\/volume %). The electrode type and unit selection capability and the ISE calibration in up to five calibration standard solutions make these instruments very useful for a large range of concentration solutions measurements.\u003c\/p\u003e\n\u003cp style=\"text-align: justify;\"\u003e \u003c\/p\u003e\n\u003cp style=\"text-align: justify;\"\u003e\u003cstrong\u003eOrder Information:\u003c\/strong\u003e HI 3222-02 (230V) are supplied with HI 1131B pH electrode. HI 7662-T temperature probe. HI 76404N electrode holder. HI 70004 pH 4.01 buffer solution sachet. HI 70007 pH 7.01 buffer solution sachet. HI 700661 electrode cleaning solution sachet (2). HI 7071S electrolyte solution (30 mL). 12 VDC adapter and instructions.\u003c\/p\u003e\n\u003cdiv\u003e\u003cstrong\u003eProduct Manuals\u003c\/strong\u003e\u003c\/div\u003e\n\u003cp style=\"text-align: justify;\"\u003e\u003cimg src=\"http:\/\/www.hannainst.com\/usa\/images\/icon-pdf.gif\" alt=\"\" width=\"20\" height=\"22\" align=\"absmiddle\" border=\"0\"\u003e Manual: \u003ca title=\"Download Product Manual\" href=\"http:\/\/www.hannainst.com\/manuals\/manHI_3220_3221_3222.pdf\" target=\"_blank\" rel=\"noopener noreferrer\"\u003eDownload\u003c\/a\u003e\u003c\/p\u003e\n\u003cp\u003e\u003cstrong\u003eSpecification :\u003c\/strong\u003e\u003c\/p\u003e\n\u003ctable class=\"spec_table\"\u003e\n\u003ctbody\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eRange\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003epH\u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e-2.0 to 20.0; -2.00 to 20.00; -2.000 to 20.000 pH\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003emV\u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e±2000 mV\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003eISE\u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e1.00 E-7 to 9.99 E10 concentration (choice of units)\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003eTemperature\u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e-20.0 to 120.0 °C (-4.0 to 248.0°F)\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eResolution\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003epH\u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e0.1; 0.01; 0.001 pH\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003emV\u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e0.1 mV\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003eISE\u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e3 digits 0.01; 0.1; 1; 10 concentration\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003eTemperature\u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e0.1°C (0.1°F)\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eAccuracy \u003csmall\u003e(@ 20ºC\/68ºF)\u003c\/small\u003e\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003epH\u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e±0.01; ±0.002 pH\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003emV\u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e±0.2 mV\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003eISE\u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e±0.5% of reading (monovalent ions). ±1% of reading (divalent ions)\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003eTemperature\u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e±0.2°C (±0.4°F) (excluding probe error)\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eCalibration\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003epH\u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003eup to five point calibration. seven standard buffers available (1.68. 4.01. 6.86. 7.01. 9.18. 10.01. 12.45) + five custom buffers\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003eISE\u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003eup to five point calibration. six standard solutions (in units selected)\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003eSlope\u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003efrom 80 to 110%\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003epH Calibration Check™\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003eyes\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eRel. mV Offset Range\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e±2000 mV\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003epH Temp. Compensation\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003emanual or automatic from -20.0 to 120.0°C (-4.0 to 248.0°F)\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eInput Channels\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e2\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003epH Electrode\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003eHI 1131B pH electrode with glass body. BNC connector and 1 m (3.3’) cable (included)\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eTemperature Probe\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003eHI 7662-T temperature probe. stainless steel with 1 m (3.3’) cable (included)\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eLogging\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003elog on demand 400 samples\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eLot Logging\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e5. 10. 30 seconds; 1. 2. 5. 10. 15. 30. 60. 120. 180 minutes (max 600 samples)\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003ePC Connectivity\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003eopto-isolated USB (with optional HI 92000 software)\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eInput Impedance\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e10\u003csup\u003e12\u003c\/sup\u003e Ohms\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003ePower Supply\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e12 VDC adapter (included)\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eEnvironment\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e0-50°C (32 to 122°F) RH max 55% non-condensing\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eDimensions\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e235 x 207 x 110 mm (9.2 x 8.14 x 4.33”)\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eWeight\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e1.8 kg (4.1 lbs.)\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003c\/tbody\u003e\n\u003c\/table\u003e\n\u003cp\u003e\u003cstrong\u003e \u003c\/strong\u003e\u003c\/p\u003e"}