{"id":1582144133,"title":"HANNA HI 96725C Kit inclding HANNA HI96725C photometer. hard carrying case","handle":"hi96725c","description":"\u003cp\u003e\u003cstrong\u003eHI 96725C\u003c\/strong\u003e\u003cspan\u003e includes HI 96725 photometer. hard carrying case. 2 sample cuvettes. scissors. cuvette cleaning cloth. battery and instruction manual.\u003c\/span\u003e\u003c\/p\u003e\n\u003cdiv id=\"Product-Body\"\u003e\n\u003cul\u003e\n\u003cli\u003eCAL CHECK™\u003c\/li\u003e\n\u003cli\u003eUser calibration\u003c\/li\u003e\n\u003cli\u003eCertified calibration and verification standards\u003c\/li\u003e\n\u003cli\u003eBEPS (Battery Error Prevention System)\u003c\/li\u003e\n\u003cli\u003eTIMER function\u003c\/li\u003e\n\u003cli\u003eAuto shut-off\u003c\/li\u003e\n\u003cli\u003eGLP Features\u003c\/li\u003e\n\u003c\/ul\u003e\n\u003cp\u003e \u003c\/p\u003e\n\u003cp\u003eLegionella species is the agent that causes human Legionnaires' disease as well as the lesser form. Pontiac Fever. Transmission is facilitated by the inhalation of mist droplets containing the Legionella bacteria.\u003c\/p\u003e\n\u003cp\u003eCommon sources of Legionella include cooling towers used in industrial cooling water systems as well as in large central air conditioning systems. domestic hot water systems. fountains. and similar disseminators that draw from a public water supply. Natural sources include freshwater ponds and creeks.\u003c\/p\u003e\n\u003cp\u003eSince Legionella is especially harmful to people with weakened immune systems. it should be actively checked for in the water systems of hospitals and nursing homes.\u003cbr\u003eThe HI 96725 measures 4 parameters that are crucial in monitoring for preventive maintenance or disinfection.\u003c\/p\u003e\n\u003cp\u003e \u003c\/p\u003e\n\u003cp\u003e \u003c\/p\u003e\n\u003cp\u003e\u003cstrong\u003eOrder Information:\u003c\/strong\u003e\u003cbr\u003e\u003cstrong\u003eHI 96725\u003c\/strong\u003e is supplied with sample cuvettes (2) with caps. 9V battery and instruction manual. \u003cbr\u003e\u003cstrong\u003eHI 96725C\u003c\/strong\u003e includes HI 96725 photometer. hard carrying case. 2 sample cuvettes. scissors. cuvette cleaning cloth. battery and instruction manual.\u003c\/p\u003e\n\u003c\/div\u003e\n\u003cdiv id=\"TabbedPanels1\" class=\"TabbedPanels\"\u003e\n\u003cul class=\"TabbedPanelsTabGroup\"\u003e\n\u003cli class=\"TabbedPanelsTab TabbedPanelsTabSelected\"\u003eSpecifications\u003c\/li\u003e\n\u003cli class=\"TabbedPanelsTab\"\u003eAccessories\u003c\/li\u003e\n\u003cli class=\"TabbedPanelsTab\"\u003eDownloads\u003c\/li\u003e\n\u003c\/ul\u003e\n\u003cdiv class=\"TabbedPanelsContentGroup\"\u003e\n\u003cdiv class=\"TabbedPanelsContent TabbedPanelsContentVisible\"\u003e\n\u003ctable class=\"spec_table\"\u003e\n\u003ctbody\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eRange\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e\u003cstrong\u003e\u003csmall\u003eCl. Free\u003c\/small\u003e\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e0.00 to 5.00 mg\/L (ppm)\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003cstrong\u003e\u003csmall\u003eCl. Total\u003c\/small\u003e\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e0.00 to 5.00 mg\/L (ppm)\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003cstrong\u003e\u003csmall\u003eCYA\u003c\/small\u003e\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e0 to 80 mg\/L (ppm)\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003cstrong\u003e\u003csmall\u003epH\u003c\/small\u003e\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e6.5 to 8.5 pH\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eResolution\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e\u003cstrong\u003e\u003csmall\u003eCl. Free\u003c\/small\u003e\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e0.01 mg\/L\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003cstrong\u003e\u003csmall\u003eCl. Total\u003c\/small\u003e\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e0.01 mg\/L\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003cstrong\u003e\u003csmall\u003eCYA\u003c\/small\u003e\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e1 mg\/L (ppm)\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003cstrong\u003e\u003csmall\u003epH\u003c\/small\u003e\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e0.1 pH\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eAccuracy @25°C\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e\u003cstrong\u003e\u003csmall\u003eCl. Free\u003c\/small\u003e\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e±0.03 mg\/L ±3% of reading\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003cstrong\u003e\u003csmall\u003eCl. Total\u003c\/small\u003e\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e±0.03 mg\/L ±3% of reading\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003cstrong\u003e\u003csmall\u003eCYA\u003c\/small\u003e\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e±1 mg\/L ±15% of reading\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003cstrong\u003e\u003csmall\u003epH\u003c\/small\u003e\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e±0.1 pH\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eLight Source\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003etungsten lamp\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eLight Detector\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003esilicon photocell with narrow band interference filter @ 525 nm\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003ePower Supply\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e9V battery\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eAuto-off\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003eafter ten minutes of non-use in measurement mode; after one hour of non-use in calibration mode; with last reading reminder\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eEnvironment\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e0 to 50°C (32 to 122°F); RH max 95% non-condensing\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eDimensions\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e192 x 104 x 69 mm (7.6 x 4.1 x 2.7\")\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eWeight\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e360 g (12.7 oz.)\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eMethod\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e\u003cstrong\u003e\u003csmall\u003eChlorine\u003c\/small\u003e\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003eadaptation of the EPA recommended DPD method 330.5\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003cstrong\u003e\u003csmall\u003eCYA\u003c\/small\u003e\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003eadaptation of the Turbidimetric method\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003cstrong\u003e\u003csmall\u003epH\u003c\/small\u003e\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003ePhenol Red method\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003c\/tbody\u003e\n\u003c\/table\u003e\n\u003cdiv\u003e\u003cstrong\u003eProduct Manuals\u003c\/strong\u003e\u003c\/div\u003e\n\u003cp\u003e\u003cimg src=\"http:\/\/www.hannainst.com\/usa\/images\/icon-pdf.gif\" alt=\"\" width=\"20\" height=\"22\" align=\"absmiddle\" border=\"0\"\u003e\u003cspan\u003e  Manual: \u003c\/span\u003e\u003ca title=\"Download Product Manual\" href=\"http:\/\/www.hannainst.com\/manuals\/man96725_10_12_10.pdf\" target=\"_blank\"\u003eDownload\u003c\/a\u003e\u003c\/p\u003e\n\u003c\/div\u003e\n\u003c\/div\u003e\n\u003c\/div\u003e","published_at":"2026-06-25T16:19:26+12:00","created_at":"2015-09-01T15:13:25+12:00","vendor":"Hanna Instruments","type":"Electrochemical","tags":["Manufacturer_Hanna Instruments","Type_Photometer"],"price":82080,"price_min":82080,"price_max":82080,"available":true,"price_varies":false,"compare_at_price":null,"compare_at_price_min":0,"compare_at_price_max":0,"compare_at_price_varies":false,"variants":[{"id":4857254981,"title":"Default Title","option1":"Default Title","option2":null,"option3":null,"sku":"HI96725C","requires_shipping":false,"taxable":true,"featured_image":null,"available":true,"name":"HANNA HI 96725C Kit inclding HANNA HI96725C photometer. hard carrying case","public_title":null,"options":["Default Title"],"price":82080,"weight":0,"compare_at_price":null,"inventory_quantity":1,"inventory_management":null,"inventory_policy":"deny","barcode":null,"requires_selling_plan":false,"selling_plan_allocations":[],"quantity_rule":{"min":1,"max":null,"increment":1}}],"images":["\/\/acornsci.co.nz\/cdn\/shop\/products\/photometer-freies-gesamtchlor-bis-1000-mgl-im-koffer-1376_1_2015_eef3c22e-76a8-4f7d-afc7-5990e96fb028.jpeg?v=1443495395"],"featured_image":"\/\/acornsci.co.nz\/cdn\/shop\/products\/photometer-freies-gesamtchlor-bis-1000-mgl-im-koffer-1376_1_2015_eef3c22e-76a8-4f7d-afc7-5990e96fb028.jpeg?v=1443495395","options":["Title"],"media":[{"alt":"HANNA HI 96725C Kit inclding HANNA HI96725C photometer. hard carrying case - Acorn Scientific","id":21293367395,"position":1,"preview_image":{"aspect_ratio":1.0,"height":800,"width":800,"src":"\/\/acornsci.co.nz\/cdn\/shop\/products\/photometer-freies-gesamtchlor-bis-1000-mgl-im-koffer-1376_1_2015_eef3c22e-76a8-4f7d-afc7-5990e96fb028.jpeg?v=1443495395"},"aspect_ratio":1.0,"height":800,"media_type":"image","src":"\/\/acornsci.co.nz\/cdn\/shop\/products\/photometer-freies-gesamtchlor-bis-1000-mgl-im-koffer-1376_1_2015_eef3c22e-76a8-4f7d-afc7-5990e96fb028.jpeg?v=1443495395","width":800}],"requires_selling_plan":false,"selling_plan_groups":[],"content":"\u003cp\u003e\u003cstrong\u003eHI 96725C\u003c\/strong\u003e\u003cspan\u003e includes HI 96725 photometer. hard carrying case. 2 sample cuvettes. scissors. cuvette cleaning cloth. battery and instruction manual.\u003c\/span\u003e\u003c\/p\u003e\n\u003cdiv id=\"Product-Body\"\u003e\n\u003cul\u003e\n\u003cli\u003eCAL CHECK™\u003c\/li\u003e\n\u003cli\u003eUser calibration\u003c\/li\u003e\n\u003cli\u003eCertified calibration and verification standards\u003c\/li\u003e\n\u003cli\u003eBEPS (Battery Error Prevention System)\u003c\/li\u003e\n\u003cli\u003eTIMER function\u003c\/li\u003e\n\u003cli\u003eAuto shut-off\u003c\/li\u003e\n\u003cli\u003eGLP Features\u003c\/li\u003e\n\u003c\/ul\u003e\n\u003cp\u003e \u003c\/p\u003e\n\u003cp\u003eLegionella species is the agent that causes human Legionnaires' disease as well as the lesser form. Pontiac Fever. Transmission is facilitated by the inhalation of mist droplets containing the Legionella bacteria.\u003c\/p\u003e\n\u003cp\u003eCommon sources of Legionella include cooling towers used in industrial cooling water systems as well as in large central air conditioning systems. domestic hot water systems. fountains. and similar disseminators that draw from a public water supply. Natural sources include freshwater ponds and creeks.\u003c\/p\u003e\n\u003cp\u003eSince Legionella is especially harmful to people with weakened immune systems. it should be actively checked for in the water systems of hospitals and nursing homes.\u003cbr\u003eThe HI 96725 measures 4 parameters that are crucial in monitoring for preventive maintenance or disinfection.\u003c\/p\u003e\n\u003cp\u003e \u003c\/p\u003e\n\u003cp\u003e \u003c\/p\u003e\n\u003cp\u003e\u003cstrong\u003eOrder Information:\u003c\/strong\u003e\u003cbr\u003e\u003cstrong\u003eHI 96725\u003c\/strong\u003e is supplied with sample cuvettes (2) with caps. 9V battery and instruction manual. \u003cbr\u003e\u003cstrong\u003eHI 96725C\u003c\/strong\u003e includes HI 96725 photometer. hard carrying case. 2 sample cuvettes. scissors. cuvette cleaning cloth. battery and instruction manual.\u003c\/p\u003e\n\u003c\/div\u003e\n\u003cdiv id=\"TabbedPanels1\" class=\"TabbedPanels\"\u003e\n\u003cul class=\"TabbedPanelsTabGroup\"\u003e\n\u003cli class=\"TabbedPanelsTab TabbedPanelsTabSelected\"\u003eSpecifications\u003c\/li\u003e\n\u003cli class=\"TabbedPanelsTab\"\u003eAccessories\u003c\/li\u003e\n\u003cli class=\"TabbedPanelsTab\"\u003eDownloads\u003c\/li\u003e\n\u003c\/ul\u003e\n\u003cdiv class=\"TabbedPanelsContentGroup\"\u003e\n\u003cdiv class=\"TabbedPanelsContent TabbedPanelsContentVisible\"\u003e\n\u003ctable class=\"spec_table\"\u003e\n\u003ctbody\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eRange\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e\u003cstrong\u003e\u003csmall\u003eCl. Free\u003c\/small\u003e\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e0.00 to 5.00 mg\/L (ppm)\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003cstrong\u003e\u003csmall\u003eCl. Total\u003c\/small\u003e\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e0.00 to 5.00 mg\/L (ppm)\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003cstrong\u003e\u003csmall\u003eCYA\u003c\/small\u003e\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e0 to 80 mg\/L (ppm)\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003cstrong\u003e\u003csmall\u003epH\u003c\/small\u003e\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e6.5 to 8.5 pH\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eResolution\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e\u003cstrong\u003e\u003csmall\u003eCl. Free\u003c\/small\u003e\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e0.01 mg\/L\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003cstrong\u003e\u003csmall\u003eCl. Total\u003c\/small\u003e\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e0.01 mg\/L\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003cstrong\u003e\u003csmall\u003eCYA\u003c\/small\u003e\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e1 mg\/L (ppm)\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003cstrong\u003e\u003csmall\u003epH\u003c\/small\u003e\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e0.1 pH\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eAccuracy @25°C\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e\u003cstrong\u003e\u003csmall\u003eCl. Free\u003c\/small\u003e\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e±0.03 mg\/L ±3% of reading\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003cstrong\u003e\u003csmall\u003eCl. Total\u003c\/small\u003e\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e±0.03 mg\/L ±3% of reading\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003cstrong\u003e\u003csmall\u003eCYA\u003c\/small\u003e\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e±1 mg\/L ±15% of reading\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003cstrong\u003e\u003csmall\u003epH\u003c\/small\u003e\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e±0.1 pH\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eLight Source\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003etungsten lamp\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eLight Detector\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003esilicon photocell with narrow band interference filter @ 525 nm\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003ePower Supply\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e9V battery\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eAuto-off\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003eafter ten minutes of non-use in measurement mode; after one hour of non-use in calibration mode; with last reading reminder\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eEnvironment\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e0 to 50°C (32 to 122°F); RH max 95% non-condensing\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eDimensions\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e192 x 104 x 69 mm (7.6 x 4.1 x 2.7\")\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eWeight\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e360 g (12.7 oz.)\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eMethod\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e\u003cstrong\u003e\u003csmall\u003eChlorine\u003c\/small\u003e\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003eadaptation of the EPA recommended DPD method 330.5\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003cstrong\u003e\u003csmall\u003eCYA\u003c\/small\u003e\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003eadaptation of the Turbidimetric method\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003cstrong\u003e\u003csmall\u003epH\u003c\/small\u003e\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003ePhenol Red method\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003c\/tbody\u003e\n\u003c\/table\u003e\n\u003cdiv\u003e\u003cstrong\u003eProduct Manuals\u003c\/strong\u003e\u003c\/div\u003e\n\u003cp\u003e\u003cimg src=\"http:\/\/www.hannainst.com\/usa\/images\/icon-pdf.gif\" alt=\"\" width=\"20\" height=\"22\" align=\"absmiddle\" border=\"0\"\u003e\u003cspan\u003e  Manual: \u003c\/span\u003e\u003ca title=\"Download Product Manual\" href=\"http:\/\/www.hannainst.com\/manuals\/man96725_10_12_10.pdf\" target=\"_blank\"\u003eDownload\u003c\/a\u003e\u003c\/p\u003e\n\u003c\/div\u003e\n\u003c\/div\u003e\n\u003c\/div\u003e"}

HANNA HI 96725C Kit inclding HANNA HI96725C photometer. hard carrying case

Product Description

HI 96725C includes HI 96725 photometer. hard carrying case. 2 sample cuvettes. scissors. cuvette cleaning cloth. battery and instruction manual.

  • CAL CHECK™
  • User calibration
  • Certified calibration and verification standards
  • BEPS (Battery Error Prevention System)
  • TIMER function
  • Auto shut-off
  • GLP Features

 

Legionella species is the agent that causes human Legionnaires' disease as well as the lesser form. Pontiac Fever. Transmission is facilitated by the inhalation of mist droplets containing the Legionella bacteria.

Common sources of Legionella include cooling towers used in industrial cooling water systems as well as in large central air conditioning systems. domestic hot water systems. fountains. and similar disseminators that draw from a public water supply. Natural sources include freshwater ponds and creeks.

Since Legionella is especially harmful to people with weakened immune systems. it should be actively checked for in the water systems of hospitals and nursing homes.
The HI 96725 measures 4 parameters that are crucial in monitoring for preventive maintenance or disinfection.

 

 

Order Information:
HI 96725 is supplied with sample cuvettes (2) with caps. 9V battery and instruction manual. 
HI 96725C includes HI 96725 photometer. hard carrying case. 2 sample cuvettes. scissors. cuvette cleaning cloth. battery and instruction manual.

  • Specifications
  • Accessories
  • Downloads
Range Cl. Free 0.00 to 5.00 mg/L (ppm)
  Cl. Total 0.00 to 5.00 mg/L (ppm)
  CYA 0 to 80 mg/L (ppm)
  pH 6.5 to 8.5 pH
Resolution Cl. Free 0.01 mg/L
  Cl. Total 0.01 mg/L
  CYA 1 mg/L (ppm)
  pH 0.1 pH
Accuracy @25°C Cl. Free ±0.03 mg/L ±3% of reading
  Cl. Total ±0.03 mg/L ±3% of reading
  CYA ±1 mg/L ±15% of reading
  pH ±0.1 pH
Light Source   tungsten lamp
Light Detector   silicon photocell with narrow band interference filter @ 525 nm
Power Supply   9V battery
Auto-off   after ten minutes of non-use in measurement mode; after one hour of non-use in calibration mode; with last reading reminder
Environment   0 to 50°C (32 to 122°F); RH max 95% non-condensing
Dimensions   192 x 104 x 69 mm (7.6 x 4.1 x 2.7")
Weight   360 g (12.7 oz.)
Method Chlorine adaptation of the EPA recommended DPD method 330.5
  CYA adaptation of the Turbidimetric method
  pH Phenol Red method
Product Manuals

  Manual: Download