{"id":2012160837,"title":"HI 98703-02 Turbidimeter US EPA with PC Interface","handle":"hi-98703-02","description":"\u003cul\u003e\n\u003cli style=\"text-align: justify;\"\u003eHigh accuracy at low ranges (below 0.05 NTU)\u003c\/li\u003e\n\u003cli style=\"text-align: justify;\"\u003eTwo. three or four point calibration\u003c\/li\u003e\n\u003cli style=\"text-align: justify;\"\u003eUSB and RS232 PC connectivity\u003c\/li\u003e\n\u003cli style=\"text-align: justify;\"\u003eLog up to 200 records\u003c\/li\u003e\n\u003cli style=\"text-align: justify;\"\u003eGLP Features\u003c\/li\u003e\n\u003cli style=\"text-align: justify;\"\u003eUser friendly. backlit  display with guidance codes\u003c\/li\u003e\n\u003cli style=\"text-align: justify;\"\u003eBattery % on startup\u003c\/li\u003e\n\u003cli style=\"text-align: justify;\"\u003eContinuous current time on display\u003c\/li\u003e\n\u003c\/ul\u003e\n\u003cp style=\"text-align: justify;\"\u003e \u003c\/p\u003e\n\u003cp style=\"text-align: justify;\"\u003e\u003cstrong\u003eThe HI 98703\u003c\/strong\u003e meets and exceeds the requirements of the USEPA Method 180.1 for wastewater and Standard Method 2130 B for drinking water. The instrument has an EPA compliance reading mode which rounds readings to meet EPA reporting requirements. Users will appreciate the accuracy and sensitivity of this instrument. particularly at very low turbidity levels.\u003c\/p\u003e\n\u003cp style=\"text-align: justify;\"\u003eThis instrument incorporates complete GLP (Good Laboratory Practice) functions that allow traceability of the calibration conditions. The last calibration points. time and date can be checked at the touch of a button.\u003c\/p\u003e\n\u003cp style=\"text-align: justify;\"\u003eWith it’s logging function. up to 200 measurements along with it’s tagged locations can be stored in internal memory and consulted at any time. Data can be later transferred to a PC via RS232 or USB interface and HANNA HI 92000 software (optional).\u003c\/p\u003e\n\u003cp style=\"text-align: justify;\"\u003e\u003cstrong\u003eExclusive Fast Tracker™\u003c\/strong\u003e\u003c\/p\u003e\n\u003cp style=\"text-align: justify;\"\u003eFor advanced field applications. the HI 98703 is equipped with Fast Tracker™—Tag Identification System (T.I.S.) that makes data collecting and management simpler than ever. Fast Tracker™ allows users to record the time and location of a specific measurement or series of measurements.\u003c\/p\u003e\n\u003cp style=\"text-align: justify;\"\u003e \u003c\/p\u003e\n\u003cp style=\"text-align: justify;\"\u003e \u003c\/p\u003e\n\u003cp style=\"text-align: justify;\"\u003e\u003cstrong\u003eOrder Information:\u003c\/strong\u003e   HI 98703-02 (230V) is supplied with sample cuvettes and caps (5). HI 98703-11 calibration cuvettes. HI 93703-58 silicone oil. cuvette cleaning cloth. batteries. AC adapter. instruction manual and rugged carrying case.\u003c\/p\u003e\n\u003cdiv\u003e\u003cstrong\u003eProduct Manuals\u003c\/strong\u003e\u003c\/div\u003e\n\u003cp style=\"text-align: justify;\"\u003e\u003cimg width=\"20\" height=\"22\" align=\"absmiddle\" alt=\"\" src=\"http:\/\/www.hannainst.com\/usa\/images\/icon-pdf.gif\" border=\"0\"\u003e  Manual: \u003ca title=\"Download Product Manual\" href=\"http:\/\/www.hannainst.com\/manuals\/manHI_98703.pdf\" target=\"_blank\" rel=\"noopener noreferrer\"\u003eDownload\u003c\/a\u003e\u003c\/p\u003e\n\u003cp\u003e\u003cstrong\u003eSpecification :\u003c\/strong\u003e\u003c\/p\u003e\n\u003ctable class=\"spec_table\"\u003e\n\u003ctbody\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eRange\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e0.00 to 9.99; 10.0 to 99.9 and 100 to 1000 NTU\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eRange Selection\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003eautomatic\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eResolution\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e0.01 NTU from 0.00 to 9.99 NTU; 0.1 NTU from 10.0 to 99.9 NTU; 1 NTU from 100 to 1000 NTU\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eAccuracy @25°C\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e±2% of reading plus 0.02 NTU\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eRepeatability\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e±1% of reading or 0.02 NTU. whichever is greater\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eStray Light\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e\u0026lt; 0.02 NTU\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eLight Detector\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003esilicon photocell\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eLight Source\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003etungsten filament lamp\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eLamp Life\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003egreater than 100.000 readings\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eMethod\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003eRatio Nephelometric Method (90°). ratio of scattered and transmitted light;  Adaptation of the USEPA Method 180.1 and Standard Method 2130 B\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eMeasuring Mode\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003enormal. average. continuous\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eTurbidity Standards \u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e\u0026lt;0.1. 15. 100 and 750 NTU\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eCalibration\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003etwo. three or four-point calibration\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eLog Memory\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e200 records\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003ePC Connectivity\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003eUSB or RS232\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eEnvironment\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003eup to 50°C (122°F); RH max 95% non-condensing\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003ePower Supply\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e1.5V AA alkaline batteries (4) or AC adapter; auto-off after 15 minutes of non-use\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eDimensions \/ Weight\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e224 x 87 x 77 mm (8.8 x 3.4 x 3.0”) \/ 512 g (18 oz.)\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003c\/tbody\u003e\n\u003c\/table\u003e\n\u003cp\u003e\u003cstrong\u003e\u003cbr\u003e\u003c\/strong\u003e\u003cstrong\u003e\u003cbr\u003e\u003c\/strong\u003e\u003c\/p\u003e","published_at":"2017-11-23T09:20:49+13:00","created_at":"2015-09-04T15:14:13+12:00","vendor":"Hanna Instruments","type":"Electrochemical","tags":["brand_Hanna","Manufacturer_Hanna Instruments","Type_Turbidity"],"price":185000,"price_min":185000,"price_max":185000,"available":true,"price_varies":false,"compare_at_price":null,"compare_at_price_min":0,"compare_at_price_max":0,"compare_at_price_varies":false,"variants":[{"id":5764195845,"title":"Default Title","option1":"Default Title","option2":null,"option3":null,"sku":"HI 98703-02","requires_shipping":false,"taxable":true,"featured_image":null,"available":true,"name":"HI 98703-02 Turbidimeter US EPA with PC Interface","public_title":null,"options":["Default Title"],"price":185000,"weight":0,"compare_at_price":null,"inventory_quantity":1,"inventory_management":null,"inventory_policy":"deny","barcode":"","requires_selling_plan":false,"selling_plan_allocations":[]}],"images":["\/\/acornsci.co.nz\/cdn\/shop\/products\/trubungs-handmessgerat-usepa-1503_1-1_2015.jpeg?v=1443494875","\/\/acornsci.co.nz\/cdn\/shop\/products\/1101_2-1_2015_ef9ced70-130a-4b0a-a7a6-7523a29be9fe.jpeg?v=1443494875"],"featured_image":"\/\/acornsci.co.nz\/cdn\/shop\/products\/trubungs-handmessgerat-usepa-1503_1-1_2015.jpeg?v=1443494875","options":["Title"],"media":[{"alt":null,"id":25183223907,"position":1,"preview_image":{"aspect_ratio":1.0,"height":800,"width":800,"src":"\/\/acornsci.co.nz\/cdn\/shop\/products\/trubungs-handmessgerat-usepa-1503_1-1_2015.jpeg?v=1443494875"},"aspect_ratio":1.0,"height":800,"media_type":"image","src":"\/\/acornsci.co.nz\/cdn\/shop\/products\/trubungs-handmessgerat-usepa-1503_1-1_2015.jpeg?v=1443494875","width":800},{"alt":null,"id":25183256675,"position":2,"preview_image":{"aspect_ratio":1.0,"height":800,"width":800,"src":"\/\/acornsci.co.nz\/cdn\/shop\/products\/1101_2-1_2015_ef9ced70-130a-4b0a-a7a6-7523a29be9fe.jpeg?v=1443494875"},"aspect_ratio":1.0,"height":800,"media_type":"image","src":"\/\/acornsci.co.nz\/cdn\/shop\/products\/1101_2-1_2015_ef9ced70-130a-4b0a-a7a6-7523a29be9fe.jpeg?v=1443494875","width":800}],"requires_selling_plan":false,"selling_plan_groups":[],"content":"\u003cul\u003e\n\u003cli style=\"text-align: justify;\"\u003eHigh accuracy at low ranges (below 0.05 NTU)\u003c\/li\u003e\n\u003cli style=\"text-align: justify;\"\u003eTwo. three or four point calibration\u003c\/li\u003e\n\u003cli style=\"text-align: justify;\"\u003eUSB and RS232 PC connectivity\u003c\/li\u003e\n\u003cli style=\"text-align: justify;\"\u003eLog up to 200 records\u003c\/li\u003e\n\u003cli style=\"text-align: justify;\"\u003eGLP Features\u003c\/li\u003e\n\u003cli style=\"text-align: justify;\"\u003eUser friendly. backlit  display with guidance codes\u003c\/li\u003e\n\u003cli style=\"text-align: justify;\"\u003eBattery % on startup\u003c\/li\u003e\n\u003cli style=\"text-align: justify;\"\u003eContinuous current time on display\u003c\/li\u003e\n\u003c\/ul\u003e\n\u003cp style=\"text-align: justify;\"\u003e \u003c\/p\u003e\n\u003cp style=\"text-align: justify;\"\u003e\u003cstrong\u003eThe HI 98703\u003c\/strong\u003e meets and exceeds the requirements of the USEPA Method 180.1 for wastewater and Standard Method 2130 B for drinking water. The instrument has an EPA compliance reading mode which rounds readings to meet EPA reporting requirements. Users will appreciate the accuracy and sensitivity of this instrument. particularly at very low turbidity levels.\u003c\/p\u003e\n\u003cp style=\"text-align: justify;\"\u003eThis instrument incorporates complete GLP (Good Laboratory Practice) functions that allow traceability of the calibration conditions. The last calibration points. time and date can be checked at the touch of a button.\u003c\/p\u003e\n\u003cp style=\"text-align: justify;\"\u003eWith it’s logging function. up to 200 measurements along with it’s tagged locations can be stored in internal memory and consulted at any time. Data can be later transferred to a PC via RS232 or USB interface and HANNA HI 92000 software (optional).\u003c\/p\u003e\n\u003cp style=\"text-align: justify;\"\u003e\u003cstrong\u003eExclusive Fast Tracker™\u003c\/strong\u003e\u003c\/p\u003e\n\u003cp style=\"text-align: justify;\"\u003eFor advanced field applications. the HI 98703 is equipped with Fast Tracker™—Tag Identification System (T.I.S.) that makes data collecting and management simpler than ever. Fast Tracker™ allows users to record the time and location of a specific measurement or series of measurements.\u003c\/p\u003e\n\u003cp style=\"text-align: justify;\"\u003e \u003c\/p\u003e\n\u003cp style=\"text-align: justify;\"\u003e \u003c\/p\u003e\n\u003cp style=\"text-align: justify;\"\u003e\u003cstrong\u003eOrder Information:\u003c\/strong\u003e   HI 98703-02 (230V) is supplied with sample cuvettes and caps (5). HI 98703-11 calibration cuvettes. HI 93703-58 silicone oil. cuvette cleaning cloth. batteries. AC adapter. instruction manual and rugged carrying case.\u003c\/p\u003e\n\u003cdiv\u003e\u003cstrong\u003eProduct Manuals\u003c\/strong\u003e\u003c\/div\u003e\n\u003cp style=\"text-align: justify;\"\u003e\u003cimg width=\"20\" height=\"22\" align=\"absmiddle\" alt=\"\" src=\"http:\/\/www.hannainst.com\/usa\/images\/icon-pdf.gif\" border=\"0\"\u003e  Manual: \u003ca title=\"Download Product Manual\" href=\"http:\/\/www.hannainst.com\/manuals\/manHI_98703.pdf\" target=\"_blank\" rel=\"noopener noreferrer\"\u003eDownload\u003c\/a\u003e\u003c\/p\u003e\n\u003cp\u003e\u003cstrong\u003eSpecification :\u003c\/strong\u003e\u003c\/p\u003e\n\u003ctable class=\"spec_table\"\u003e\n\u003ctbody\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eRange\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e0.00 to 9.99; 10.0 to 99.9 and 100 to 1000 NTU\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eRange Selection\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003eautomatic\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eResolution\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e0.01 NTU from 0.00 to 9.99 NTU; 0.1 NTU from 10.0 to 99.9 NTU; 1 NTU from 100 to 1000 NTU\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eAccuracy @25°C\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e±2% of reading plus 0.02 NTU\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eRepeatability\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e±1% of reading or 0.02 NTU. whichever is greater\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eStray Light\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e\u0026lt; 0.02 NTU\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eLight Detector\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003esilicon photocell\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eLight Source\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003etungsten filament lamp\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eLamp Life\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003egreater than 100.000 readings\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eMethod\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003eRatio Nephelometric Method (90°). ratio of scattered and transmitted light;  Adaptation of the USEPA Method 180.1 and Standard Method 2130 B\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eMeasuring Mode\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003enormal. average. continuous\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eTurbidity Standards \u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e\u0026lt;0.1. 15. 100 and 750 NTU\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eCalibration\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003etwo. three or four-point calibration\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eLog Memory\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e200 records\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003ePC Connectivity\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003eUSB or RS232\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eEnvironment\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003eup to 50°C (122°F); RH max 95% non-condensing\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003ePower Supply\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e1.5V AA alkaline batteries (4) or AC adapter; auto-off after 15 minutes of non-use\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003ctr\u003e\n\u003ctd\u003e\u003cstrong\u003eDimensions \/ Weight\u003c\/strong\u003e\u003c\/td\u003e\n\u003ctd\u003e \u003c\/td\u003e\n\u003ctd\u003e\u003csmall\u003e224 x 87 x 77 mm (8.8 x 3.4 x 3.0”) \/ 512 g (18 oz.)\u003c\/small\u003e\u003c\/td\u003e\n\u003c\/tr\u003e\n\u003c\/tbody\u003e\n\u003c\/table\u003e\n\u003cp\u003e\u003cstrong\u003e\u003cbr\u003e\u003c\/strong\u003e\u003cstrong\u003e\u003cbr\u003e\u003c\/strong\u003e\u003c\/p\u003e"}

HI 98703-02 Turbidimeter US EPA with PC Interface

Product Description
  • High accuracy at low ranges (below 0.05 NTU)
  • Two. three or four point calibration
  • USB and RS232 PC connectivity
  • Log up to 200 records
  • GLP Features
  • User friendly. backlit  display with guidance codes
  • Battery % on startup
  • Continuous current time on display

 

The HI 98703 meets and exceeds the requirements of the USEPA Method 180.1 for wastewater and Standard Method 2130 B for drinking water. The instrument has an EPA compliance reading mode which rounds readings to meet EPA reporting requirements. Users will appreciate the accuracy and sensitivity of this instrument. particularly at very low turbidity levels.

This instrument incorporates complete GLP (Good Laboratory Practice) functions that allow traceability of the calibration conditions. The last calibration points. time and date can be checked at the touch of a button.

With it’s logging function. up to 200 measurements along with it’s tagged locations can be stored in internal memory and consulted at any time. Data can be later transferred to a PC via RS232 or USB interface and HANNA HI 92000 software (optional).

Exclusive Fast Tracker™

For advanced field applications. the HI 98703 is equipped with Fast Tracker™—Tag Identification System (T.I.S.) that makes data collecting and management simpler than ever. Fast Tracker™ allows users to record the time and location of a specific measurement or series of measurements.

 

 

Order Information:   HI 98703-02 (230V) is supplied with sample cuvettes and caps (5). HI 98703-11 calibration cuvettes. HI 93703-58 silicone oil. cuvette cleaning cloth. batteries. AC adapter. instruction manual and rugged carrying case.

Product Manuals

  Manual: Download

Specification :

Range   0.00 to 9.99; 10.0 to 99.9 and 100 to 1000 NTU
Range Selection   automatic
Resolution   0.01 NTU from 0.00 to 9.99 NTU; 0.1 NTU from 10.0 to 99.9 NTU; 1 NTU from 100 to 1000 NTU
Accuracy @25°C   ±2% of reading plus 0.02 NTU
Repeatability   ±1% of reading or 0.02 NTU. whichever is greater
Stray Light   < 0.02 NTU
Light Detector   silicon photocell
Light Source   tungsten filament lamp
Lamp Life   greater than 100.000 readings
Method   Ratio Nephelometric Method (90°). ratio of scattered and transmitted light;  Adaptation of the USEPA Method 180.1 and Standard Method 2130 B
Measuring Mode   normal. average. continuous
Turbidity Standards   <0.1. 15. 100 and 750 NTU
Calibration   two. three or four-point calibration
Log Memory   200 records
PC Connectivity   USB or RS232
Environment   up to 50°C (122°F); RH max 95% non-condensing
Power Supply   1.5V AA alkaline batteries (4) or AC adapter; auto-off after 15 minutes of non-use
Dimensions / Weight   224 x 87 x 77 mm (8.8 x 3.4 x 3.0”) / 512 g (18 oz.)